USP5 Transdermal Patch Holder, 35mm 40 Mesh. Outer diameter is 41.2mm and meets the diagram in the USP
OEM Part# 12-4200
Transdermal Patch Holder with 35mm (1.38") opening and 40 Mesh screen.
Use in Paddle Over Disk Applications
Transdermal or patch testing is carried out using USP method 5 (paddle over disc) or USP method 6, the rotating cylinder.
With paddle over disc, the transdermal patch is placed between a glass disc and an inert PTFE mesh. This is placed at the bottom of the vessel, with the mesh facing upwards, under a rotating paddle.
Unlike dissolution testing, transdermal testing is carried out at 32°C to reflect the lower temperature of the skin. Other variables such as the height setting and sampling requirements are the same as dissolution testing. APPFIVE-V35 has an outer diameter of 41.2mm and meets the diagram in the USP.
Length | 1.5 inch |
---|---|
Pack Size | Each |
Transdermal | USP5 Disc |